SARS-CoV-2 Total Antibody Calibration Procedure
SARS-CoV-2 Total Antibody Calibration Procedure
Siemens Dimension Exl SARS Co V2 Total antibody calibration procedure Video The Dimension Exl SARS Co V2 total antibody assay is for use in the qualitative detection of total antibodies to SARS Co V2 prior to processing the Dimension Exl CV2 TSA for patient samples. QC or calibration allow all other tests to complete. Ensure that no other essays are submitted for processing until all SARS Co V2 total antibody testing is complete and the analyzer is in standby. From the operating menu, select F1, Enter Data and enter the following calibrator information. In the position field, enter the segment letter and position number of Cal 2 for patient name. Enter the calibrator name for sample number. Enter the calibrator lot number. Leave the location field blank. Tests Request 3 tests for the dimension EXL, CV2, TSA mode. Select F7 next mode to change to sample cup priority. Select F4 next priority to change to QC fluid. Select F8 next fluid to change to serum QC 3. Note when processing dimension EXL CV2T, serum QC 3 is reserved for the calibrator. An erroneous calibration of the assay could occur if any other fluid designation or material is used. Next, add a minimum of 0.1 milliliters of the SARS Co V2 Total Antibody Calibrator Level 2 to a 1.5 milliliter sample cup. Place the sample cup into the segment position entered in the position field and select F2 Process Single. The following message will appear. This fluid serum QC 3 will calibrate ACV 2T method. Continue yes or no. Select Y to begin the calibration. When the calibration is complete, run the SARS Co V2 total antibody quality control.
SARS-CoV-2 Total Antibody Calibration Procedure Video SIEMENS Standby Sampler Idle IMT OK 05/27/20 05:04 Healthineers Dimension Exit labmanager en Enter OPERATING MENU STAT Status PLEASE SELECT AN OPTION USING THE FUNCTION KEYS: Sample Alert At any time, press HELP to get help, or press EXIT to exit a program. Supplies QC Alert Calib Alert To Shut down or Change Operator, press Sign Out or Alt+EXIT. Run Home F1: ENTER DATA F2: SAMPLE STATUS F3: TEST RESULTS F4: SYSTEM PREP F5: PROCESS CTRL F6: SYSTEM CONFIG F7: DIAGNOSTICS F8: SEGMENT STATUS ENTER SAMPLE DATA Position: Mode: PRIMARY TUBE Patient Name: Priority: ROUTINE Sample No .: Fluid: SerumQC1 Location: Dilution: Tests: Volume: ul F1: NEW SAMPLE F2: PROCESS SINGLE F3: LOAD LIST F4: NEXT PRIORITY F5: DELETE TEST F6: DELETE SAMPLE F7: NEXT MODE F8: NEXT FLUID A1 CV2T Cal Patient Name: CV2T Cal 1234AB CV2T 10 ul CV2T CV2T CV2T CV2T CV2T Volume: 30 ul 30 ul Mode: SAMPLE CUP Priority: QC This fluid (SerumQC3) will calibrate a CV2T method. Continue? (y/n) YES NO Disclaimer Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively "Operator Manual") issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively "Functionality") may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright @ Siemens Healthcare GmbH 2020 91052 Erlangen, GermanylTelephone: +49 9131 84-Olsiemens-healthineers.com
- SARS-CoV-2 Total Antibody Calibration Procedure