POCcelerator™ Web View and Manage Linearity Results Job Aid
This job aid provides an overview of POCcelerator™ Web's three linearity result view types and instructions to add comments to linearity results, resolve incomplete linearity measurements, and record the review of linearity results.
POCcelerator™ Web – SIEMENS Healthineers View and Manage Linearity Results Job Aid View Linearity Results The Linearity monitor provides three views for linearity Status Messuremen ... Location Device Device model Lot number Expiration date | Activation date|Comments Reviewer Reviewed Resolve Q Q Q Q Q Q results: Lo 8/16/2019 LOC1 00001507192 StatStrip Lot 100 1/1/2020 8/26/2019 4:06. All linearity results: displays sample results 8/28/2019 LOC1 00001507192 StatStrip Lot100 1/1/2020 8/26/2019 4:06 ... • received from a device using the linearity lot in a Status Analyte Sample Unit Slope Intercept Correlation given day group in a single linearity result Glu_mmol Vollblut mmol/l 1 • Analyte detail view: displays the analytes that were measured with calculated slope, intercept and Status Measurement time Level Reagent Value Flag Low High Target Comments Linearity Results Report 3:25:34 AM 1 1001 12 10 20 15 correlation values 3:25:34 AM 2 1001 80 50 90 70 800 Analyte measurements view: displays the analyte 3:25:34 AM 3 |1001 138 130 150 140 • 3:25:34 AM 4 1001 222 200 250 225 600 measurement time per lot level with sample value 3:25:34 AM 5 1001 705 700 710 705 and target ranges 10 20 50 Page 1 of 1 (5 entries) < 1 > 400 200 Measured Value Note: Users must have the appropriate permissions 0 to the Web Linearity Monitor user group to view 200 O 007 600 Target Value and manage linearity lots. Range Target Passed Trend Failed 1 Effective Date: 20 Apr 2022 | HOOD05162003270754 Unrestricted © Siemens Healthineers, 2022 POCcelerator™ Web – SIEMENS Healthineers View and Manage Linearity Results Job Aid Add Comments to Linearity Results Linearity monitor When adding a comment to a linearity result with errors, the status will change from red to yellow. -Range of time Device model Technical errors --- - 1. Select Linearity > Linearity monitor Month 3/12/2022 Until 4/12/2022 Analytes 2. Select the appropriate filter options Re-check Record Review Lab Comment 3. Select one or more linearity results and select Export Reset filters Lab Comment All linearity results Incomplete linearity measurements Review linearity results 4. Enter a lab comment Status Measurem ... Y Location Y Devic Y Device model Lot number Expiration ... Y Activation ... Comments Reviewer Reviewed Y Resolve Q Q 5. Select Save Q to No data 10 20 50 Page 1 of 1 (0 entries) 1 Copyright @ 2022 Siemens Healthcare Diagnostics 2 Effective Date: 20 Apr 2022 | HOOD05162003270754 Unrestricted © Siemens Healthineers, 2022 POCcelerator™ Web – SIEMENS Healthineers View and Manage Linearity Results Job Aid Resolve Incomplete Linearity Measurements 1. Select Linearity > Linearity monitor 2. Select the appropriate filter options Linearity monitor 3. Select the Incomplete linearity measurements tab 2 Range of time Device model Technical errors 4. Hover over the red Status button of an incomplete Apply Analytes 8/31/2017 Until 2/7/2010 result to view the error 3 7 8 Re-check Record Review Lab Comment 5. Select Details for more information All linearity results Incomplete linearity measurements Review linearity results 6. Resolve the issue Status Measurement time ocation Y |Device Y Device model User Lot number Expiration date Level Reagent Exclude 7. On the Linearity monitor screen, select Re-check Q Q Q O O O 10/1/2018 6:27 AM ICU 3 - MISC Analyzers 78957 Accuchek inform II 1000 1234 10/31/2018 10004R Exclude Details W • The result status will change from red to green 4 10/1/2018 6:28 AM ICU 3 - MISC Analyzers 78957 Accuchek inform II 1000 1234 10/31/2018 10004R Exclude Details 5 after all result errors have been resolved 10/1/2018 6:29 AM ICU 3 - MISC Analyzers 78957 Accuchek inform II 1000 1234 10/31/2018 10004R Exclude Details W 8. Select the result line and select Lab Comment 10 20 50 Page 1 of 1 (3 items) < 1 > 9. Enter a comment Copyright @ 2018 Siemens Healthcare Diagnostics Inc. 10. Select Save 3 Effective Date: 20 Apr 2022 | HOOD05162003270754 Unrestricted © Siemens Healthineers, 2022 POCcelerator™ Web – SIEMENS Healthineers View and Manage Linearity Results Job Aid Record the Review of Linearity Results 1. Select Linearity > Linearity monitor 2. Select the appropriate filter options 3. Select the Review linearity results tab 2 Linearity monitor Range of time Device model Technical errors ---- 4. Select one or more linearity results from the ---- Apply Analytes 8/31/2017 Until 2/7/2019 table and select Record Review 4 3 Record Review Lab Comment 5. Select Yes to confirm All linearity results Incomplete linearity measurements Review linearity results Reset filters The result can be viewed under the All Status Measurement date Y |Location Y Device Y Device model Lot number Expiration date Y Activation date Comments Reviewer Reviewed • Q Q Q Q Q Q O linearity results tab or by selecting the 10/1/2018 ICU 3 - MISC Analyzers 78957 Accuchek inform II 1234 10/31/2018 10/1/2018 2:08 PM green status icon 10 20 50 Page 1 of 1 (1 items) < 1 V • The result is updated with the Reviewer Copyright @ 2018 Siemens Healthcare Diagnostics Inc. name and Reviewed date and time v4.18.16.3 4 Effective Date: 20 Apr 2022 | HOOD05162003270754 Unrestricted © Siemens Healthineers, 2022 SIEMENS Healthineers Please note that the learning material is for training purposes only. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Please note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. POCcelerator and all associated marks are trademarks of Siemens Healthcare Diagnostics Inc., or its affiliates. All other trademarks are the property of their respective owners. © Siemens Healthcare GmbH 2022 Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens-healthineers.com 5 Effective Date: 20 Apr 2022 | HOOD05162003270754 Unrestricted © Siemens Healthineers, 2022
- poccelerator
- poccelerator web
- pocweb
- linearity
- results
- record
- comment
- incomplete